Radiometric image quality is expressed in terms of :
Radiometric calibration is used to assign a physical meaning to the digital value of each pixel, by relating it to a reference value called radiance at the top of the amosphere (expressed in W.m-2.sr-1.mm-1) by means of an absolute calibration coefficient.
The value of this coefficient, which is specific for each spectral band, can be estimated by several methods using :
A combination of all these methods is used to obtain the calibration coefficient at a given time, to monitor changes over time and to determine the calibration accuracy. For VGT the calibration accuracy is estimated to be 8% at the beginning of life and 5% after a few months.
Radiometric resolution represents the smallest radiance variation which the instrument can detect. It is estimated by measuring different values of noise on the images (noise along columns, on a digital quick-look, on the complete image) while the instrument is observing a uniform area.
The modulation transfer function characterises the sharpness of the image and is obtained through spectral analysis of simultaneous HRVIR and VGT images.
Radiometric performance is expressed at the system level in terms of :
* the column noise (NeDLD)
which is the radiance variation at the entrance to the instrument, corresponding to the
standard deviation measured along a column of the raw image data,
* the digital quick-look noise (NeDLH) which is the radiance variation at the entrance to the
instrument, corresponding to the standard deviation measured on a digital quick-look
normalized to 60 x 60 pixels,
* the image noise (NeDLS)
which is the radiance variation at the entrance to the instrument, corresponding to the
standard deviation measured on an image normalized to 1 728 x 2 000 pixels. As this
characteristics cannot be directly measured, it is estimated by combining 2 noises (column
and normalization).
Radiometric performance characteristics :
Spectral band | Useful observable Radiance (L2) W.m-2.sr-1.micron-1 |
NeDLD | NeDLH | NeDLS |
B0 | 11 | 0.13 | 0.30 | 0.53 |
B2 | 110 | 0.10 | 0.40 | 0.87 |
B3 | 106 | 0.12 | 0.28 | 0.39 |
SWIR | 20 | 0.02 | 0.05 | 0.085 |
* of the absolute calibration coefficient used to convert the instrument's output data into physical values (radiance at the top of the atmosphere), is referred to as absolute calibration,
* of the ratio of absolute calibration coefficients for the HRVIR and VGT instruments for a given spectral band, is known as HRVIR/VGT inter-instrument calibration.
Calibration accuracy :
Calibration accuracy | Absolute | Inter-instrument HRVIR/VGT |
after 6 months in orbit | 5 to 8% | 5% |
Geometric accuracy :
Field of view depending on the camera | Parallelism with ground tracks | Line distortion | Registration |
101.01° to 101.08° | <45 m rd | < 70 millipixels | < 100 millipixels |
page updated on the 05-07-04